Khan, Muhammad Tahir (1969) Diagnostic Accuracy of Ultrasound B scan using 10 MHz linear probe in ocular trauma; results from a high burden country. Pakistan Journal of Medical Sciences, 32 (2). ISSN 1681-715X
9241-44195-1-PB.pdf - Published Version
Download (743kB)
Abstract
Objective: To study the diagnostic accuracy of Ultrasound B scan using 10 MHz linear probe in ocular trauma.
Methods: A total of 61 patients with 63 ocular injuries were assessed during July 2013 to January 2014. All patients were referred to the department of Radiology from Emergency Room since adequate clinical assessment of the fundus was impossible because of the presence of opaque ocular media. Based on radiological diagnosis, the patients were provided treatment (surgical or medical). Clinical diagnosis was confirmed during surgical procedures or clinical follow-up.
Results: A total of 63 ocular injuries were examined in 61 patients. The overall sensitivity was 91.5%, Specificity was 98.87%, Positive predictive value was 87.62 and Negative predictive value was 99%.
Conclusion: Ultrasound B-scan is a sensitive, non invasive and rapid way of assessing intraocular damage caused by blunt or penetrating eye injuries.